Nitrogen-phosphorus detector (NPD) definition

 A  B  C  D  E  F  G  H  I  J  K  L  M  N  O  P  Q  R  S  T  U  V  W  X  Y  Z

 


The overall design of a nitrogen-phosphorus detector (NPD) is similar to a flame-ionization detector (FID). The major difference is that the hydrogen/air flame of the FID is replaced by a heated rubidium silicate bead in the NPD. The effluent from the GC column passes through the hot bead. The hot rubidium salt emits ions when nitrogen and phosphorus-containing compounds pass over it. The ions are collected on a collector above the heated bead to produce a current, similar to the FID.

Schematic of NPD See the FID document.